Abbott Medical (CRM-Sylmar) Aveir¿ Leadless Pacemaker, Right Atrium Back to Search Results
Model Number LSP201A
Medical Device Problem Codes Failure to Interrogate (1332); Stretched (1601); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Embolism/Embolus (4438)
Date of Event 07/27/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that routine post-implant procedure diagnostic imaging was performed and dislodgment of the right atrial (ra) leadless pacemaker (lp) into the pulmonary artery was observed.The ra lp was unable to be interrogated.A revision procedure was performed and the ra lp was successfully retrieved.Helix damage was noted on the ra lp, suspected to have occurred during the retrieval.The ra lp was replaced.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Aveir¿ Leadless Pacemaker, Right Atrium
Common Device Name
Leadless Pacemaker
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50210233
Report Number2017865-2026-16821
Device Sequence Number21334485
Product Code PNJ
UDI-Device Identifier05415067040701
UDI-Public(01)05415067040701(10)S000108365(17)271010
Combination Product (Y/N)Y
Initial Reporter CountryEZ
PMA/510(K) Number
P150035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSP201A
Device Lot NumberS000108365
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-