| Model Number |
TRIEEA28MT |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Tissue Breakdown (2681)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic right hemicolectomy procedure, in performing the anastomosis between the transverse colon and right colon.Poor staple formation was noted in approximately one-quarter of the staple, with excessive tissue incorporated.To resolve the issue, the anastomosis was reinforced with sutures.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|