• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-ANGIO DRAPE CARDIAC PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-ANGIO DRAPE CARDIAC PACK Back to Search Results
Catalog Number DYNJ51615P
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that (b)(6) is experiencing packs with damaged contrast squeeze lines and kinked/damaged waste bag lines.There has also been damage to the guidewire bowl.
 
Additional Manufacturer Narrative
It was reported that the customer is experiencing packs with damaged contrast squeeze lines and kinked/damaged waste bag lines.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-ANGIO DRAPE CARDIAC PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
diana arizmendi
three lakes drive
northfield, IL 60093--275
MDR Report Key50210292
Report Number1417592-2026-15408
Device Sequence Number14977033
Product Code OEZ
UDI-Device Identifier10198459135767
UDI-Public10198459135767
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ51615P
Device Lot Number25AMA031
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/03/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2025
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-