• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-MINISTRY C SECTION PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-MINISTRY C SECTION PACK Back to Search Results
Catalog Number DYNJ82919
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "we had a gown in our c-section pack have significant strike through this morning.".
 
Additional Manufacturer Narrative
According to the customer, while using a "level 3 gown" from the c section pack on (b)(6) 2023, the surgeon had "blood" strikethrough the gown onto her "arm".No sample or photo was received for the investigation; finished good production records were reviewed and no non-conformances were noted that would have contributed to the reported issue, and the finished good pack build was reviewed with no issues that might have an impact on the issue.The reported problem could not be confirmed because no sample or photo was received; the cause was documented as dc: traced to user knowledge.The surgical gowns currently used for the pack are classified as aami pb70 barrier level 3 rated and are designed to be fluid resistant, however with enough force it may allow strikethrough to occur; based on these details, the item performed to specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-MINISTRY C SECTION PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
juan mota
three lakes drive
northfield, IL 60093--275
MDR Report Key50210562
Report Number1417592-2026-15422
Device Sequence Number17672702
Product Code OHM
UDI-Device Identifier40195327219278
UDI-Public40195327219278
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ82919
Device Lot Number23EBB359
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-