According to the customer, while using a "level 3 gown" from the c section pack on (b)(6) 2023, the surgeon had "blood" strikethrough the gown onto her "arm".No sample or photo was received for the investigation; finished good production records were reviewed and no non-conformances were noted that would have contributed to the reported issue, and the finished good pack build was reviewed with no issues that might have an impact on the issue.The reported problem could not be confirmed because no sample or photo was received; the cause was documented as dc: traced to user knowledge.The surgical gowns currently used for the pack are classified as aami pb70 barrier level 3 rated and are designed to be fluid resistant, however with enough force it may allow strikethrough to occur; based on these details, the item performed to specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|