| Model Number |
380620-48 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Thrombosis/Thrombus (4440); Liver Failure (4492); Metabolic Acidosis (4923)
|
| Date of Event |
07/11/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that a patient who underwent a da vinci-assisted hepatic lobectomy procedure, developed post-hepatectomy liver failure and expired.The patient was child-pugh class a.There were no intraoperative complications, and the procedure was successfully completed robotically.The surgery involved parenchymal resection, described as approximately 10% non-anatomical resection.The suspected cause of deterioration was portal vein thrombosis that progressed to ischemic hepatitis, accompanied by refractory acidosis and a rising international normalized ratio (inr).Despite intensive care unit level support and other unspecified interventions, the patient expired on postoperative day 2.Attempts were made to obtain additional information from the surgeon; however, no additional information was received.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the intraoperative system logs found the system functioned normally with no errors logged and no indications relating to this event.Log review of the 5 subsequent procedures found the system functioned normally; no intraoperative events or issues found.The following concomitant products were used during this procedure: 30 degree endoscope plus, monopolar curved scissors, large needle driver, cadiere forceps, and mbf.A site history review found no complaints were made for the instruments used during this procedure.A review of the event performed by an intuitive surgical medical safety officer (mso) concluded that the patient in this report died following a hepatic resection secondary to portal vein thrombosis in the setting of coagulopathy, acidosis, and class a cirrhosis.How this portal vein thrombosis occurred is not provided.No intraoperative complications were reported.Based on the information provided in the summary of events, insufficient information is available to determine if any intuitive surgical products or instruments contributed to this event.
|
| |
|
Search Alerts/Recalls
|
|
|