Model Number NEU_INS_STIMULATOR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Skin Erosion (2075); Therapeutic Effects, Unexpected (2099)
Event Date 07/23/2015
Event Type  Injury  
Manufacturer Narrative
Concomitant product: product id: 977a290, serial# (b)(4), implanted: (b)(6) 2015, explanted: (b)(6) 2015, product type: lead.(b)(4).
 
Event Description
The health care provider (hcp) reported via the company representative (rep) that on (b)(6) 2015 the lead, implanted in the occipital right, was removed in emergency after "the externalization of the electrode under the skin." in absence of the electrode, the pain felt by the patient was not correctly/properly relieved.The issue was resolved at the time of the report.If additional information is received, a follow up report will be sent.
 
Manufacturer Narrative
Device analysis for lead (b)(4) revealed no significant anomaly.The lead body was cut through, product segmented.The lead was cut 77.5cm from proximal end.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE NEUROSTIMULATOR
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5021186
MDR Text Key23696373
Report Number3007566237-2015-02344
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Followup
Report Date 05/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNEU_INS_STIMULATOR
Device Catalogue NumberNEU_INS_STIMULATOR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/21/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/03/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00052 YR
-
-