Baxter received a report that golvo 9000 had power cable and control box burnt.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
|
The baxter technician found the control box, battery, charging cable and hand control needed to be replaced.The golvo mobile lift in intended to be used for transferring patients (adult or children) between devices (e.G., within the room), floor lifting, horizontal lifting, supporting patient limbs, ambulating patient, bathing patient, toileting patient, weighing patient and transferring patients from car.Intended for use in following environments: health care, intensive care, emergency ward, rehabilitation, habilitation.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the control box, battery, charging cable and hand control to resolve the reported event.Based on this information, no further action is required.
|