| Model Number |
60005 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Visual Impairment (2138); Diplopia/Double Vision (4802)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A site reported to rxsight that a bilaterally implanted patient's light adjustable lens was explanted from the right eye due to a shift in refraction and monocular diplopia after completion of light treatments.A new lal was implanted as a replacement.Postoperatively, the site reported that the patient was satisfied with the results of the lens exchange and that the diplopia in the right eye was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing records were reviewed and found no deviations, non-conformances, or other manufacturing issues related to the reported event.A review of light delivery device data revealed that the final lock-in light treatment was delivered to an inadequately dilated eye, which could have contributed to the patient's symptoms.Manufacturer reference #: (b)(4).
|
| |
|
Search Alerts/Recalls
|