Model Number 60005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Diplopia/Double Vision (4802)
Date of Event 07/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A site reported to rxsight that a bilaterally implanted patient's light adjustable lens was explanted from the left eye due to a shift in refraction and monocular diplopia after completion of light treatments.A new lal was implanted as a replacement.No additional information is available.
 
Additional Manufacturer Narrative
Manufacturing records were reviewed and found no deviations, non-conformances, or other manufacturing issues related to the reported event.A review of light delivery device data revealed that the final lock-in light treatment was delivered to an inadequately dilated eye, which could have contributed to the patient's symptoms.Manufacturer reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Light Adjustable Lens (LAL)
Common Device Name
Light Adjustable Lens (LAL)
Manufacturer (Section D)
RxSight, Inc.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RxSight, Inc.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
jim zhang
100 columbia
aliso viejo, CA 92656
+1(949)273-6163
MDR Report Key50212626
Report Number3012712027-2026-00379
Device Sequence Number9915667
Product Code PZK
UDI-Device Identifier00818806020685
UDI-Public(01)00818806020685(11)230821(17)260731
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL07-005100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-