Model Number 72404133
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 07/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence.A surgical procedure was performed during which the cuff was downsized, and the pump was replaced and repositioned.There were no further patient complications reported.
 
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence.A surgical procedure was performed during which the cuff was downsized, and the pump was replaced and repositioned.There were no further patient complications reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100857486.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100735900.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of "urinary incontinence" was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of adverse event related to patient condition was assigned to this investigation, which means the patients condition, disease, or anatomy is demonstrably responsible for the adverse event and use of the device has neither caused nor otherwise influenced the adverse event.Model number/catalog number: 72400024 serial number: n/a batch/lot number: 1100857486 model/catalog description: balloon fgs 61-70 cm unique identifier (udi) # (b)(4) model number/catalog number: 72404127 serial number: n/a batch/lot number: 1100735900 model/catalog description: control pump fgs iz unique identifier (udi) # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50213561
Report Number2124215-2026-42958
Device Sequence Number17666784
Product Code EZY
UDI-Device Identifier00878953003115
UDI-Public(01)00878953003115(17)271115(10)1100832202
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404133
Device Catalogue Number72404133
Device Lot Number1100832202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2026
Supplement Date Received by Manufacturer08/24/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date09/16/2026
Date Device Manufactured11/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexUnknown
-
-