The device was not returned to stryker sustainability solutions for evaluation.The submitted medsun form stated that the device was not available for evaluation.As the device was not returned for evaluation, inspection was unable to be performed.As the complaint device was not returned, a conclusive root cause could not be determined, and the reported event could not be confirmed.A review of the dhr for the reported lot/serial number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: excessive force.Contact of needle point with hard object.Shipping/handling damage or extreme conditions.Attempting to bend or otherwise alter the shape of the device/shaft.The instructions for use (ifu) state: "before beginning the procedure, verify overall compatibility of all instruments and accessories." "inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery." the reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|