| Model Number |
40700BAXUS |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a novum iq large volume pump turned on; however, there was no display output.This was observed in the biomedical service department.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.During functional testing, the display issue was identified; the pump powered on and no content was displayed.A device history review revealed no issues that could have caused or contributed to the reported issue.The reported condition was verified.The cause of the condition was found to be a defective display.The front panel assembly would be replaced to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|