ABBOTT VASCULAR INC. HI-TORQUE BALANCE MIDDLEWEIGHT¿; Wire, Guide, Catheter Back to Search Results
Model Number 1009664
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a heavily tortuous right radial artery.A hi-torque versaturn guide wire and ht bmw universal ii guide wire were introduced and advanced through the tortuous segment.It was noted that the bmw guide wire met resistance with the anatomy during advancement.Upon removal, resistance was noted due to the bmw guide wire winding, and force was applied.The tip separated; therefore, the right femoral artery was successfully accessed, and a 6f sheath was placed.A snare device was advanced through the sheath, and the separated tip was successfully retrieved.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other similar incidents from this lot.Reportedly, during advancement, resistance was noted due to anatomy and the bmw guide wire core separated.Upon removal, resistance was noted and force was applied.It should be noted the hi-torque balance middleweight universal ii guide wire instructions for use (ifu) states: do not: push, auger, withdraw, or torque a guide wire that meets resistance.The force applied during removal of the device seemed to be a reasonable clinical response to the difficulties.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that resistance was met with the heavily tortuous anatomy resulting in the reported difficult to advance and ultimately resulted in the reported core material separation.During removal of the compromised device interaction with the heavily tortuous anatomy resulted in the reported difficult to remove.As reported, a snare device was advanced through the sheath and successfully retrieved the separated bmw guide wire.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.H6: medical device problem code 2017 - excessive force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT¿
Common Device Name
Wire, Guide, Catheter
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50214288
Report Number2024168-2026-03142
Device Sequence Number13261690
Product Code DQX
UDI-Device Identifier08717648118715
UDI-Public(01)08717648118715(11)250529(17)270430(10)5052974
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152709
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1009664
Device Catalogue Number1009664
Device Lot Number5052974
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HT VersaTurn guidewire
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-