BOSTON SCIENTIFIC CORPORATION Maverick?; catheters, transluminal coronary angioplasty, percutaneous
|
Back to Search Results |
|
| Model Number |
H7493892815250 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 90% stenosed target lesion was located in the mildly tortuous and severely calcified proximal to mid right coronary artery.A 2.50mm x 15mm maverick? balloon catheter was used for pre-dilatation at 8 to 12 atmospheres.However, during the pre-dilatation, the proximal bend first and the balloon ruptured.The procedure was completed and another device was used to complete the procedure.Intraoperative ecg and hemodynamic parameters remained stable, and the patient experienced no specific discomfort.He was returned to the ward postoperatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.With all the available information, boston scientific concludes:
device analysis findings:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review:
review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review:
a risk review was performed and confirmed that the event was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
this investigation is assigned a most probable investigation conclusion code of adverse event related to patient condition as it is most likely that anatomical factors (90% stenosed target lesion was located in the mildly tortuous and severely calcified proximal to mid right coronary artery) were met which resulted in the reported event.This code was selected as the most probable complaint cause based on the information available.Without a returned device it is not possible to confirm the complaint allegation of balloon rupture.
|
| |
|