| Model Number |
353-01-106 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
06/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Evaluation:\
the agent reported \"patient had surgery done 6 weeks ago, press fit tibia collapsed from fracture.Complete revision.\" the previous surgery and the surgery detailed in this event occurred 38 days apart.\
item number: 353-01-106.\
item description: djo empowr kneetm, press fit bp, 6l.\
lot number: 191w1093.\
part revision: d.\
product type: knee.\
manufacture date: 3/18/24.\
expiration date: 3/12/30.\
the reported device was returned to surgical and examination confirmed; the pegs are sheared, pegs not returned.\
this evaluation is limited in scope as the item(s) associated with this investigation was not returned to (b)(6) for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.\
there was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.\
there were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.\
root cause:\
the root cause of this complaint was a revision surgery due to press fit tibia collapsed from fracture.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.\
\
containment:\
inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.\
\
device history records review:\
a review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.\
\
complaint history:\
customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.This is the 1st complaint against the part/lot number combination.
|
| |
|
Event or Problem Description
|
|
Patient had surgery done 6 weeks ago ((b)(6) 2024); press fit tibia collapsed from fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation:\
the agent reported \"patient had surgery done 6 weeks ago, press fit tibia collapsed from fracture.Complete revision.\" the previous surgery and the surgery detailed in this event occurred 38 days apart.\
item number: 353-01-106\
item description: djo empowr kneetm, press fit bp, 6l\
lot number: 191w1093\
part revision: d\
product type: knee\
manufacture date: 3/18/24\
expiration date: 3/12/30\
this evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.\
there was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.\
there were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.\.\
root cause:\
the root cause of this complaint was a revision surgery due to press fit tibia collapsed from fracture.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical\
\
containment:\
inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.\
\
device history records review:\
a review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.\
\
complaint history:\
customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.This is the 1st complaint against the part/lot number combination.
|
| |
|
Event or Problem Description
|
|
Patient had surgery done 6 weeks ago ((b)(6) 2024), press fit tibia collapsed from fracture.
|
| |
|