|
During a gastrointestinal dilation procedure, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.It was reported that there was a small hole in the balloon as it would not inflate to (b)(6) "[atm]".A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation evaluation:
a product evaluation was performed only by the picture provided in response to this report because the product said to be involved was not provided to cook for evaluation.Per the photo provided we cannot complete a full evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on statements describing the event.The picture shows the label on the pouch, and the lot number matches the report.The device is not visible in the picture provided.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion:
we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The report states that lubrication was not applied to the balloon prior to advancing down the endoscope accessory channel.Applying lubrication will aid in balloon preservation and optimize balloon performance.The instructions for use direct the user to "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel".This activity will aid in endoscopic advancement and balloon preservation.The additional information provided also states that it is unknown if the user applied negative pressure to the balloon prior to advancement through the endoscope accessory channel.A possible contributing factor to a leakage is failure to apply negative pressure to the balloon prior to advancement.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use also advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." this activity will also aid in endoscopic advancement and balloon preservation.A leakage/damage to the balloon material can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Another possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage wire guided balloons esophageal-pyloric-colonic are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action:
based on the quality engineering risk assessment no corrective action is warranted at this.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.Additional comments:
based on the information provided that lubrication was not applied to the balloon prior to advancement through the endoscope, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|