| Model Number |
380677-31R |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to starting a da vinci-assisted surgical procedure using an xi system, and after surgical port placement, an intuitive surgical, inc.(isi) field service engineer (fse) called from offsite to report error 903 occurred.The technical support engineer (tse) reviewed the logs and found error 25514 for remote arm controller (rac) 2.The site performed a couple of power cycles, but the error persisted.The procedure was converted to another dv system, and no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Fse was able to reproduce the error issue with remote arm controller (rac) and replaced it to correct the reported problem.System was tested and verified as ready for use.The complaint was confirmed based on the field evaluation.Isi has not received the product associated with this event to perform failure analysis.Therefore, the root cause of the customer reported failure mode has not been determined.
|
| |
|