| Model Number |
470067-13 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during central processing with a 30-degree endoscope plus, a piece on endoscope that attaches to the robot for use was separated from the viewing piece of the scope.There was no report of patient involvement or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The endoscope was analyzed and found with a dislodged component.The camera instrument was found dislodged, the ring and adapter was return.As an additional observation not reported by site, the endoscope was visually inspected and found with damage to the button pack assembly.Visual inspection confirmed hairline crack(s) on l/r of the button pack assembly.The endoscope was visually inspected and was found with damage at the distal end.The distal window located at distal tip was visually inspected and found cracked.The endoscope was visually inspected and glue damage on fiber distal tip (gap) was found.During inspection of the endoscope distal tip, the fiber damage.The endoscope cable integrated connector was visually inspected and found with scratches at trim plate.The cable integrated connector exhibited scratch marks.The endoscope cable integrated connector was visually inspected and found with mechanical damage.The cable integrated connector exhibited mechanical damage such as scratch marks at cable connector proximal end.The complaint was confirmed by failure analysis.The endoscope was placed through friction test using a screening aid to manually rotate the adapter to detect for friction.The camera instrument adapter did not rotate properly and/or noises were heard during testing and confirmed as binding.The probable root cause of this binding is attributed to component susceptibility to increased friction.
|
| |
|
Search Alerts/Recalls
|