| Model Number |
470184-16 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The permanent cautery spatula was analyzed and found to have a dislodged crimp from the yaw pulley.The yaw cable crimp is installed.Visual inspection observed the yaw cable crimp is not properly seated within the yaw pulley as the spatula moves in yaw.The visual inspection revealed no signs of missing parts.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.
|
| |
|