| Model Number |
IO-401 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
Dyspnea (1816)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's device was displaying the system error message and was not providing enough oxygen.As a result, the patient experienced shortness of breath.The patient did not have a backup at the time of the event.The patient has since received their replacement device and is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
The unit was received with a reported system error, which was confirmed during testing.The unit shut down after a few minutes due to the compressor failure.Because compressor harness was damaged.Furthermore, the psa cycle (according to the data log) being high (14) is an indication that the zeolite is getting contaminated by moisture.Uip replaced due to cosmetic damage.
|
| |
|