Model Number IO-401
Medical Device Problem Code Insufficient Flow or Under Infusion (2182)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 07/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's device was displaying the system error message and was not providing enough oxygen.As a result, the patient experienced shortness of breath.The patient did not have a backup at the time of the event.The patient has since received their replacement device and is doing well.
 
Additional Manufacturer Narrative
The unit was received with a reported system error, which was confirmed during testing.The unit shut down after a few minutes due to the compressor failure.Because compressor harness was damaged.Furthermore, the psa cycle (according to the data log) being high (14) is an indication that the zeolite is getting contaminated by moisture.Uip replaced due to cosmetic damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen Rove 4 Portable Oxygen Concentrator
Common Device Name
Inogen Rove 4 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50216564
Report Number3004672275-2026-00245
Device Sequence Number6099695
Product Code CAW
UDI-Device Identifier00817131020414
UDI-Public(01)00817131020414
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K222086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-401
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/31/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-