It was reported that prior to starting a da vinci-assisted surgical procedure using a xi system, an operating room (or) staff called in to report that the right high resolution stereo viewer (hrsv-r) on the surgeon side console (ssc) was black during case setup.The intuitive surgical, inc.(isi) technical support engineer (tse) walked the customer through a hard power cycle, but there was no change.As a result, the surgeon elected to abort to another dv system.Upon review, the tse found error 121 on ssc 2, pointing to a suspect right eye hrsv monitor.No known impact or patient consequence was reported.
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse replaced two high resolution stereo viewer (hrsv) and personality module surgeon console (pmsc) to resolve the issue.The system was verified and ready to use.Isi has not received the devices for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.
|