Model Number URF-P6
Medical Device Problem Codes Material Fragmentation (1261); Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the uretero-reno fiberscope to the service center for a leak and image issue which did not indicate an mdr reportable event.During the evaluation of the device, it was found that the adhesive was lifted and chipped.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.The root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Common Device Name
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50217444
Report Number9610595-2026-65074
Device Sequence Number21481804
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172298
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberURF-P6
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/11/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date08/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-