• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic Back to Search Results
Model Number 72400098
Medical Device Problem Codes Malposition of Device (2616); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this artificial urinary sphincter (aus) pump was mispositioned in the patient's scrotum.It was believed that due to the patient's size, placing the pump lower would allow the patient to better access the pump.The patient later reported difficulty accessing the pump due to its location.As a result, the pump was explanted and replaced.No patient complications were reported.
 
Event or Problem Description
It was reported that this artificial urinary sphincter (aus) pump was mispositioned in the patient's scrotum.It was believed that due to the patient's size, placing the pump lower would allow the patient to better access the pump.The patient later reported difficulty accessing the pump due to its location.As a result, the pump was explanted and replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72400024, serial number: n/a, batch/lot number: 1100687795, model/catalog description: balloon fgs 61-70 cm, unique identifier (udi) #: (b)(4).Model number/catalog number: 72400160, serial number: n/a, batch/lot number: 1100817998, model/catalog description: belt cuff fgs 4.0 cm, unique identifier (udi) #: (b)(4).
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this artificial urinary sphincter (aus) underwent a thorough analysis.The pump was visually and microscopically inspected, and a leakage and functional tests were performed.As a result of these tests, no anomalies or damage were identified.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications prior to distribution.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of difficult to press and malposition of device was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, cause not established was assigned to this investigation.Model number/catalog number: 72400024, batch/lot number: 1100687795, model/catalog description: balloon fgs 61-70 cm, unique identifier (udi) #: (b)(4).Model number/catalog number: 72400160, batch/lot number: 1100817998, model/catalog description: belt cuff fgs 4.0 cm, unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50217668
Report Number2124215-2026-43115
Device Sequence Number10584635
Product Code EZY
UDI-Device Identifier00878953000688
UDI-Public(01)00878953000688(17)301019(10)1100817899
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72400098
Device Catalogue Number72400098
Device Lot Number1100817899
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/23/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Supplement Date Received by Manufacturer08/21/2026
Initial Report FDA Received Date08/12/2026
Supplement Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
Patient SexMale
-
-