It was reported that a patient underwent a premature ventricular contraction (pvc) ablation procedure with a decanav catheter and the patient experienced a cardiac tamponade which required pericardiocentesis.While mapping the right ventricle during a pvc case, an abnormal path of the decanav catheter through the right ventricle into the possible pericardium was noticed.Physician was immediately notified of the catheter tracing and ultrasound images.After further assessment, the physician performed pericardiocentesis and 155 ml of fluid removed by syringe and an additional 20 ml was collected in a drainage system.The patient was stable.The patient required extended hospitalization due to drain placement.There were two to three sheath and catheter exchanges during the procedure.A total of 28 ablations were performed in the right ventricle.The correct catheter settings were selected on the generator.There were no issues with flow rate change at the start of ablation.No error messages were observed on biosense webster equipment during the procedure.The physician stated that there was no default in the product and the event was operator error, as she put the catheter in the wrong place.
|
|
D4: udi: the lot number provided of ad26110015 was not recognized in the system.Therefore, the full udi is currently not available.The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number:
(b)(4).
|