| Model Number |
480299-08 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during a da vinci assisted surgical procedure that a suction irrigator was unable to irrigate and use suction.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: no fragments fell into the patient.There was no concern if any fragments fell into the patient because it failed before use.The customer confirmed that no damage or missing material occurred during the surgical procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The suction irrigator instrument was analyzed and found to have missing pieces measured approximately 0.283"" x 0.173"" and were not returned with the instrument.Components adjacent to the distal tip do not show any damage.The broken distal tip caused the instrument not to irrigate as reported by the customer.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.
|
| |
|