| Model Number |
GS-100 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
Dyspnea (1816)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's unit had white dust coming out of the bottom following a column replacement.It was also reported that the unit was not producing enough oxygen for the patient.As a result, the patient had difficulty breathing and was taken to the er.The patient's oxygen needs may have changed as they decided to go with a higher output machine instead of getting a replacement device.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event is estimated.The unit was returned for evaluation on (b)(6) 2026.The return reason of oxygen error is confirmed since zeolite dust existed in the unit, which is caused by the breakdown of zeolite particles.This occurs when zeolite rubs against one another.Over time this can lead to multiple errors.This dust is so fine it can escape the column frits and contaminate the muffler with zeolite dust.
|
| |
|