| Model Number |
1000 |
| Medical Device Problem Code |
Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During annual service stryker observed that the device's defibrillation charge is delayed.In this state defibrillation therapy may be delayed or not be available if needed.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker confirmed the reported issue during annual service.The customer declined repair of the device and instead plan to replace with a new device.The device was returned unrepaired to the customer.The cause of the reported issue could not be determined.
|
| |
|