| Model Number |
U1054260 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Electrical /Electronic Property Problem (1198); Communication or Transmission Problem (2896); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The trilogy 100 ventilator was returned to the third-party service center for remediation.The device was not in use on a patient at the time of the occurrence.During the evaluation the device failed a test step for communication during testing.Upon further investigation of the device the service center states the device does not power on and has a burning smell.The device was disassembled and the power board was found to be burned.In conclusion, the system board with daughter and power management board were replaced to address the issues.The service center confirmed both power-on and communication were successfully validated after the repairs.Additionally, during the service of the device, the device failed testing for sd locked card inserted: sd detected unrelated to the reportable malfunction.The sd card socket does not retain the memory card, causing the reported failure.The system board with daughter was installed to validate the device, and it completed all tests successfully.Components were replaced as part of maintenance of the device or due to physical/cosmetic damage.The device was retested and passed all test.This device been serviced, inspected, tested, met acceptance criteria in accordance with all the applicable customer requirements as defined in the contractual agreement with plexus.The device history record has been reviewed and approved by plexus quality assurance.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure were reviewed to assess for the observed probability levels and compare with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Search Alerts/Recalls
|
|
|