It was reported that when using the bd pyxis¿ pro medstation main station displayed a black, blank screen and was completely unresponsive.The customer stated that the station remained on a black screen with no visible content.A reboot of the station was attempted; however, the issue persisted and the station did not recover.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 29-may-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the station was stuck on the black screen and totally blank.A field service engineer verified that the touchscreen was non-functional while the peripherals, scanner, and keyboard remained powered and operational, replaced the touchscreen on the pyxis pro, connected and secured all cabling, and ensured proper device alignment, powered on the system and confirmed the touchscreen was fully functional.Allowed the device to complete synchronization, then verified full system operation.The customer confirmed functionality of the cordless scanner and successfully completed inventory transactions.The system functioned as intended after the field service engineer repaired the device.
|