The 6629 bed rails were installed on a drive medical bed, model 15013-l, which is not an invacare product.The bed and rails were delivered to the patient on (b)(6) 2025.An invacare sps1080 mattress was also being used on the bed; it was delivered to the patient on (b)(6) 2026.There was no noted malfunction at the time of the event.Entrapment is a known risk when using bed rails.The 6629 bed rail assembly, installation and operating instructions warn, "entrapment may occur.Proper patient assessment and monitoring, and proper maintenance and use of equipment is required to reduce the risk of entrapment.Variations in bed rail dimensions, and mattress thickness, size or density could increase the risk of entrapment.Bed rails with dimensions different from the original equipment supplied or specified by the bed manufacturer may not be interchangeable and may result in entrapment or other injury." vitas healtchare was informed that the invacare mattress and rails may not be compatible with the drive medical bed, and it was recommended that they contact drive medical to confirm.The serial number provided for the bed rails (b)(6) is incomplete and was unable to be verified.Therefore, it is unclear when the rails were manufactured or by whom.The 6629 rails have been manufactured by invacare at invamex, and they have also been manufactured by a supplier and then distributed by invacare.Since the manufacturer cannot be confirmed, this report was filed under the invacare corporation dba invacare america registration.Should additional information become available, a supplemental record will be filed.
|