Model Number 1020007
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the everflo device.There was a report of device interior full of white powder.There was no allegation of serious or permanent harm or injury.No medical intervention was specified by the patient.The device was returned to the third-party service center for evaluation.During the evaluation of the device there was evidence of front cabinet has external burned, label warning need to be replaced, rear cabinet need to be replaced because is smell a tobacco, sieve is leakage zeolite, tube braided silicone needs to be replaced, manifold and opi concentrator tubing are out of use is yellowed and fissure and is losing air and contaminated with tobacco, flowmeter is broken, inlet filter need to be replaced for preventive maintenance.The device was requested to be sent to the manufacturer's product investigation lab (pil) for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EverFlo
Common Device Name
Generator, Oxygen, Portable
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50220977
Report Number2518422-2026-042997
Device Sequence Number10286139
Product Code CAW
UDI-Device Identifier00606959032569
UDI-Public(01)00606959032569
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 08/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1020007
Device Catalogue Number1020007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2025
Initial Report FDA Received Date08/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-