The account reported finding light handle cover breaks during removal.A photo and a sample were received for the investigation; the sample evaluation found two light handle covers, one with the inner ring completely broken off, and finished good production records, finished good pack build, trending, component trending, the bill of materials (bom), and production records were reviewed.Based on the review of the reported complaint, the provided sample and photo, and the investigation, medline confirmed the reported issue.Further investigation revealed that the root cause is attributed to a vendor product issue related to the component manufacturing process; the complaint was provided to the component manufacturing facility for further evaluation, supplier quality will monitor this issue for trending purposes, and the incident was found to be acceptable.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|