The account reported finding irrigator leaking.A sample was received; medline received one used suction irrigator for sample review, no additional pack contents were returned, there were no cracks or kinks seen with the tubing, the remaining liquid within the tubing was manipulated and it did not leak, both buttons were tested and they did not stick, and the batteries were not returned so functionality or power could not be tested.As no defects were noted with the received sample, medline could not confirm the reported defect.Finished good production records and the finished good pack build were reviewed, no issues were noted that could contribute to the reported issue, the complaint was provided to the vendor for further evaluation, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|