| Model Number |
02.12.3D05L |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Aseptic loosening (4883)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery due to 3d metal cemented tibial tray loosening after about 8 months from primary.The surgeon revised the gmk 3dmetal tibial tray size 5l to a gmk tibial tray cemented left s4 and revised the gmk-sphere tibial insert e-cross - flex 5l - 10mm to a gmk-sphere tibial insert e-cross - flex 4l - 14mm.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 29 july 2026 lot.2507354: (b)(4) items manufactured and released on 18-jun-2025.Expiration date: 2030-jun-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Root cause:implant mobilization is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|