Model Number 1201
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had high impedance.They experienced no falls or trauma to their implant site.The patient had surgery to revise their tined lead and neurostimulator (ins).During the revision, it was noted the lead was sheared off in the middle.The provider wanted to replace both the tined lead and ins.Additionally, after a one-week follow-up, the patient reported replacing a pad every time they use the bathroom due to some cognitive decline.The patient had their stimulation increased until they felt the sensation in the vaginal area.Another assessment and possible increase is scheduled in the next few weeks.There was no patient complications reported.
 
Additional Manufacturer Narrative
Block d2b: product code: additional product code qon.Udi for concomitant product, neurostimulator: block d10: model number/catalog number: 4101, serial number: (b)(6) batch/lot number: ax1t068879, model/catalog description: neurostimulator, unique identifier (udi) #: (b)(4).Block g4: premarket / 510(k) #: additional premarket / 510(k) # p190006.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Axonics
Common Device Name
Stimulator, Electrical, Implantable, For Incontinence
Manufacturer (Section D)
Axonics Modulation Technologies, Inc.
15515 sand canyon ave
irvine CA 92618
Manufacturer (Section G)
Axonics Inc
15515 sand canyon ave
irvine CA 92618
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50223046
Report Number2124215-2026-43156
Device Sequence Number16061446
Product Code EZW
UDI-Device Identifier10810005340141
UDI-Public(01)10810005340141(11)250423(17)280423
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P180046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number1201
Device Catalogue Number1201
Device Lot NumberAL10005868
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/18/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/12/2026
Supplement Report FDA Received Date09/04/2026
Date Device Manufactured04/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-