• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Entrant VR; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Entrant VR; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) Back to Search Results
Model Number CDVRA300Q
Medical Device Problem Codes Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
Following a magnetic resonance imaging (mri), the device was found to be in backup vvi mode (bvvi) resulting in high-voltage (hv) therapy being disabled.The device had not been programmed into mri mode.Technical support was contacted, and the device was reprogrammed to resolve the event.The patient was stable with no consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Entrant VR
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50223761
Report Number2017865-2026-16901
Device Sequence Number18750912
Product Code NIK
UDI-Device Identifier05415067032836
UDI-Public(01)05415067032836(10)P000254167(17)280331
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDVRA300Q
Device Catalogue NumberCDVRA300Q
Device Lot NumberP000254167
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Initial Report FDA Received Date08/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-