| Model Number |
Patient Information Center iX |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on a patient information center ix indicating that following a network failure at the hospital, the philips monitoring systems were unable to resume all functions automatically after the network was restored.The bed-to-bed overview and alarm forwarding were only reactivated automatically in the medical intensive care unit and the intermediate care unit.In all other stations with philips monitoring, this had to be manually activated in the pic ix system configuration.The device was in use on a patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by remote service personnel which included remotely checking the system configuration.The results of the analysis indicate the auto reconnect setting was deactivated on some of the control units.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was the configuration to auto reconnect was deactivated on some of the control units, causing them to remain in local mode and requiring manual reconnection.The issue was resolved by activating the auto reconnect function in configuration.A review of the service history for this device shows the problem has not been reported again for this device.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|
Search Alerts/Recalls
|