The customer observed falsely elevated alinity i free t4 results for two samples.The following data was provided (normal range 9.01 to 19.05 pmol/l):
(b)(6) 2026
patient 1 (79-year-old female)
initial result was 19.31 pmol/l, specimen was retested again within 45 min interval, and result was 16.58 / 18.07 / 17.35 pmol/l,
patient 2 (53-year-old female)
initial result was 22.96 pmol/l, specimen was retested again within 45 min interval, and result was 13.50 pmol/l.The customer recentrifuged specimen and results were 13.82 and 14.06 pmol/l.No impact to patient management was reported.
|