| Model Number |
SPL-S |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the lithotripsy system, which is an olympus asset with no reported complaint.During the evaluation of the device, faulty transducer receptacles unit at front with burn marks inside was observed.The investigator indicated that the most likely cause of the faulty transducer receptacles with burn marks was due to component failure.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|