• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Gyrus ACMI, Inc. SHOCKPULSE-SE LITHOTRIPSY SYSTEM; lithotriptor, ultrasonic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Gyrus ACMI, Inc. SHOCKPULSE-SE LITHOTRIPSY SYSTEM; lithotriptor, ultrasonic Back to Search Results
Model Number SPL-S
Medical Device Problem Codes Thermal Decomposition of Device (1071); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the lithotripsy system, which is an olympus asset with no reported complaint.During the evaluation of the device, faulty transducer receptacles unit at front with burn marks inside was observed.The investigator indicated that the most likely cause of the faulty transducer receptacles with burn marks was due to component failure.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE-SE LITHOTRIPSY SYSTEM
Common Device Name
lithotriptor, ultrasonic
Manufacturer (Section D)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50224410
Report Number3011050570-2026-00812
Device Sequence Number12142836
Product Code FEO
UDI-Device Identifier00821925044197
UDI-Public00821925044197
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142428
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberSPL-S
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/14/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/16/2026
Initial Report FDA Received Date08/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-