There was an allegation of questionable elecsys cmv igg and elecsys cmv igm results for 1 patient on a cobas e 801 analytical unit compared to a vidas analyzer.This medwatch will cover cmv igg.Refer to medwatch with a1 patient identifier (b)(6) for information on the cmv igm results.On (b)(6) 2025, the patient had a cmv igg result of 0.38 u/ml (negative) and a cmv igm result of 0.69 coi (negative).On (b)(6) 2026, the patient had a cmv igg result of 3.68 u/ml (positive) and a cmv igm result of 0.657 coi (negative).On (b)(6) 2026, the customer repeated the previous patient samples.The sample from on (b)(6) 2025 gave a cmv igg result of 0.37 u/ml (negative) and a cmv igm result of 0.789 coi (equivocal).The sample was also sent to another laboratory and tested on a vidas analyzer and the cmv igg result was 20 (positive) and the cmv igm result was 33.6 (positive).The sample from on (b)(6) 2026 gave a cmv igg result of 3.69 u/ml (positive) and a cmv igm result of 0.718 coi (equivocal).The sample was also sent to another laboratory and tested on a vidas analyzer and the cmv igg result was 19 (positive) and the cmv igm result was 24 (positive).The vidas analyzer results units of measurement were not provided.
|