The customer reported that 2 of the packs came opened and were not usable.A photo and a sample were received; in both, the edge seam of the pouch on the tyvek end was open along the top, middle of the bag, finished good production records and finished good pack build were reviewed, and finished good and component stock checks found no defects.Based on the complaint details, photo and sample evaluation, and investigation, medline confirmed the reported issue.The root cause was determined to be a vendor product issue relating to the component manufacturing process; the complaint was provided to the vendor for further evaluation, supplier quality will monitor this issue for trending purposes, and further risk assessment found the incident acceptable w/ justification.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|