It was reported that during a vein closure procedure using the micro introducer 7f 7cm 7fx7cm, a floopy part of the wire broke off inside the patient¿s vein and a portion of the sheath was left inside the patient.The event involved the right distal great saphenous vein.Local anesthesia was utilized, and a guidewire was used for insertion.The instructions for use were followed without issue, no resistance was encountered when advancing the device, and no packaging issues were noted.An x-ray was ordered to assess the retained fragment(s).The vein closure procedure was completed and the vein was reported as closed.
|
|
We conducted a thorough review of the manufacturing and inspection records for this lot and found no deviations or non-conformances.No samples were returned from the customer for review.Additionally, no photographic or visual evidence of any kind was provided, which would have allowed the allegation to be reviewed.Without the affected device, a comprehensive investigation cannot be conducted, and a root cause cannot be determined.As a result, no corrective action is required at this point.However, if samples are returned later, we would be happy to reopen this complaint for re-evaluation.
|