| Model Number |
D121 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Use of Device Problem (1670); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited a code 1006, indicative of a high voltage detected on a lead during a device charge.Technical services (ts) reviewed the available data and determined that the low-voltage code was valid, with an irregular decrease in battery voltage potentially related to an internal electrical short, noted an alert for remote monitoring disabled (rmd) on radio frequency (rf) telemetry.Subsequently, this icd was successfully explanted and replaced.No additional adverse patient effects were reported outside the revision procedure.At this time, this product has not been returned.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|