| Catalog Number |
04Z21-25 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a false positive alinity i troponin result generated on an alinity i processing module (ai34060) for a 28-year-old male patient.The patient came to the emergency room and troponin timed studies were ordered.The following data was provided:
first timed result sid 26439958 = <2.7 ng/l
second timed result sid 26440077 = <2.7 ng/l
third timed result sid 26440328 = 62.2 ng/l, repeat on another analyzer (ai34061) = <2.7 ng/l
fourth timed result (on ai 34061) sid 26440856 = <2.7 ng/l
a corrected report was issued for the 3rd elevated troponin result.The specimen was hemolyzed on the elevated troponin; however, the rerun was on the same hemolyzed sample generating the result of <2.7 ng/l.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.
|
| |
|