| Lot Number |
0036130399 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Angina (1710)
|
| Date of Event |
06/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It has been reported that exertional angina pectoris has occurred post procedure.During pulse field ablation procedure to treat paroxysmal atrial fibrillation farawave was selected for use.The patient's condition is improving.This case was reported as a worsening of the patient's pre-existing condition (other than af) and was considered related to the procedure (vascular access).The device is not expected to be returning as it has been disposed.
|
| |
|
Event or Problem Description
|
|
It has been reported that exertional angina pectoris has occurred post procedure.During pulse field ablation procedure to treat paroxysmal atrial fibrillation farawave was selected for use.The patient condition is improving.This case was reported as a worsening of the patient's pre-existing condition (other than af) and was considered related to the procedure (vascular access).The device is not expected to be returning as it has been disposed.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.Investigation summary:
with all the available information, boston scientific is unable to confirm cause of the reported clinical observation of e061201 angina.The device has not been received for analysis; therefore, a technical analysis of the complaint device could not be completed.Device history record review:
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Risk review:
a risk review performed for the farawave device confirmed that the event of e061201 angina is a known event defined in the product risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
boston scientific has assigned an investigation conclusion code of adverse event related to procedure and supporting evidence that came to this conclusion.Boston scientifics investigation findings were the exertional angina occurred post-procedure and was specifically reported as associated with the procedure/vascular access, in the setting of a worsening of the patient's pre-existing condition other than atrial fibrillation.Although the underlying condition may have contributed to the patients susceptibility to the event, the reported procedural association provides the strongest support for a procedure-related conclusion.No device-related allegations were identified for this complaint.The device met all requirements prior to being approved for final distribution/sale and there was no evidence to suggest that the instructions for use (ifu) was not followed.Correction to sections a1, g2, and h6 evaluation method codes.
|
| |
|