It was reported that a patient experienced vision blurred and numbness in the left arm.During a pulse field ablation procedure (pfa) to treat atrial fibrillation a farawave cathether and a versacross connect access solution for faradrive were selected for use.No issues with the transseptal puncture was reported.Right inferior pulmonary vein (ripv) was extremely small, while in basket, 301 errors were received, catheter was repositioned and still received errors and ultimately the catheter formed a cobra shape.It was removed the catheter and manually separated the splines, went back into the left atrium (la) and tried to reattempt basket in the ripv, catheter went cobra again.There was difficulty delivering the basket into the ripv.Bubbles appeared while delivering in basket configuration due to something that typically occurs which is caused by an inadequate tissue contact because the basket is undersized.If the electrodes are too close, the system has a built-in safety mechanism that automatically discontinues energy delivery.Physician indicated that there may have been excessive therapy delivery in the ripv region.Physician believes that 301 errors and additional application outside of workflow contributed to the issue.Catheter was replaced because it assumed a cobra shape configuration and after replacing it was segmented the ripv with flower.Procedure was completed.It was notified that patient experienced symptoms concerning for a possible transient ischemic attack (tia).The patient underwent imaging to evaluate for a stroke, which did not demonstrate evidence of a stroke.It was stated that the patient is now doing well; however, the symptoms occurred on two separated occasions during recovery and included left arm numbness and blurred vision.No further information was provided.
|