| Model Number |
D022 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Palpitations (2467); Shock from Patient Lead(s) (3162)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the health care professional (hcp) requested a review of reports for this implantable cardioverter defibrillator (icd).Technical services (ts) reviewed the data and noted the rhythm appeared to be supraventricular tachycardia (svt) or 1:1 conduction.Ts also noted multiple anti-tachycardia pacing (atp) and shocks were delivered for the suspected 1:1 rhythm.It was also reported the patient experienced palpitations.No additional adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|