Model Number D022
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes Palpitations (2467); Shock from Patient Lead(s) (3162)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the health care professional (hcp) requested a review of reports for this implantable cardioverter defibrillator (icd).Technical services (ts) reviewed the data and noted the rhythm appeared to be supraventricular tachycardia (svt) or 1:1 conduction.Ts also noted multiple anti-tachycardia pacing (atp) and shocks were delivered for the suspected 1:1 rhythm.It was also reported the patient experienced palpitations.No additional adverse patient effects were reported.This icd remains in service.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? MINI ICD DR
Common Device Name
Implantable cardioverter defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50228421
Report Number2124215-2026-43339
Device Sequence Number16016685
Product Code LWS
UDI-Device Identifier00802526533907
UDI-Public(01)00802526533907(17)211031
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2021
Device Model NumberD022
Device Catalogue NumberD022
Device Lot Number122177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date08/12/2026
Supplement Report FDA Received Date09/23/2026
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age43 YR
Patient SexFemale
-
-