| Model Number |
IO-501 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The unit received a complaint of a burning smell; the complaint was confirmed, melted outside upper housing and lower housing caused by fire sources.The unit was found to have low oxygen that is caused by column contaminated due to absorbing too much moisture.The muffler was filled with dust, which contributed to the blockage.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's device charger burned and damaged the device while it was being charged.A burning smell was observed due to the event.It was noted that there was minor property damage, but there were no reported injuries or negative impact to patient's health due to the event.This report will be on the patient's device.
|
| |
|