Event or Problem Description
|
|
It was reported that pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.The physician was attempting to go transeptal and noticed a small pericardial effusion.The physician had slipped into the coronary sinus with the pigtail catheter.The patient was given protamine and the effusion remained stable.The procedure was discontinued.The patient was monitored and will have follow up echocardiography imaging to evaluate for effusion.
|
| |
|
Event or Problem Description
|
|
It was reported that pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.The physician was attempting to go transeptal and noticed a small pericardial effusion.The physician had slipped into the coronary sinus with the pigtail catheter.The patient was given protamine and the effusion remained stable.The procedure was discontinued.The patient was monitored and will have follow up echocardiography imaging to evaluate for effusion.
|
| |
|
Event or Problem Description
|
|
It was reported that pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.The physician was attempting to go transeptal and noticed a small pericardial effusion.The physician had slipped into the coronary sinus with the pigtail catheter.The patient was given protamine and the effusion remained stable.The procedure was discontinued.The patient was monitored and will have follow up echocardiography imaging to evaluate for effusion.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete udi and other product specific information.If additional details become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: patient codes - updated.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes:
device technical analysis:
the watchman fxd curve? access system was discarded; therefore, returned device analysis could not be completed.Device history record review:
a review was completed of the device history records (dhr) for the watchman fxd curve? access system, part # m635tu80020 batch # 0038803442.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman fxd curve? access system instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include pericardial effusion and cardiac perforation.Risk review:
a risk review was completed and confirmed that the events of pericardial effusion and cardiac perforation were defined in the risk documentation.These event types have been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|