| Model Number |
430100-02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during central processing, a single port monopolar curved scissors instrument had broken tip.There was no report of patient involvement.Intuitive surgical, inc.(isi) contacted the customer and obtained the following information: the instrument was not used on the patient in the procedure, so no fragments fell into the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the single port monopolar curved scissors (mcs) instrument to perform failure analysis (fa).Fa was able to confirm/reproduce the reported complaint.The instrument was found to have a broken instrument tube adapter.This component is located at the distal end of the instrument and serves as the interface between the instrument and the user-installed tip accessory.The broken detached piece was not returned and measures approximately 1.99 mm x 2.71 mm in size.The manual articulation of inputs was performed and passed.The housing was removed for internal inspection and found to have no damage.The probable root cause of a damaged tube adapter is attributed to either tip accessory installation issues or damage during use.
|
| |
|