INTUITIVE SURGICAL, INC EndoWrist SP; MONOPOLAR CURVED SCISSORS Back to Search Results
Model Number 430100-02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during central processing, a single port monopolar curved scissors instrument had broken tip.There was no report of patient involvement.Intuitive surgical, inc.(isi) contacted the customer and obtained the following information: the instrument was not used on the patient in the procedure, so no fragments fell into the patient.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive the single port monopolar curved scissors (mcs) instrument to perform failure analysis (fa).Fa was able to confirm/reproduce the reported complaint.The instrument was found to have a broken instrument tube adapter.This component is located at the distal end of the instrument and serves as the interface between the instrument and the user-installed tip accessory.The broken detached piece was not returned and measures approximately 1.99 mm x 2.71 mm in size.The manual articulation of inputs was performed and passed.The housing was removed for internal inspection and found to have no damage.The probable root cause of a damaged tube adapter is attributed to either tip accessory installation issues or damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EndoWrist SP
Common Device Name
MONOPOLAR CURVED SCISSORS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50231275
Report Number2955842-2026-35589
Device Sequence Number13348434
Product Code NAY
UDI-Device Identifier00886874123454
UDI-Public(01)00886874123454(10)K10250717
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K241814
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number430100-02
Device Catalogue Number430100
Device Lot NumberK10250717 0018
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/23/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/17/2026
Initial Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient SexUnknown
-
-